Product Introduction
Intended Use
Kidney Function(Uric Acid)Test Strip works with the Renal Function Analysis Meter or Dry Biochemical Analysis Meter to measure the Uric Acid concentration in whole blood, plasma and serum. Kidney Function(Uric Acid)Test Strip is for professional use.
Principle and reference values
Kidney Function(Uric Acid)Test Strip use a timed-endpoint method to measure the Uric Acid (UA) concentrations in whole blood, serum or plasma. The system monitors the change in absorbance at 620 nm at a fixed-time interval. The change in absorbance is directly proportional to the concentration of metabolite in the specimen.In the reaction, the uric acid is oxidized to allantoin and hydrogen peroxide by uricase. Peroxidase catalyzes the reaction of hydrogen peroxide with 4-aminoantipyrine and phenol to produce a colored quinoneimine product. The change in absorbance is directly proportional to the concentration of uric acid in the specimen.
Reference values are listed in the chart below
Tests | Desirable |
Uric Acid(UA) | Men: 0.202~0.416mmol/L(3.40~6.99mg/dL) Women:0.142~0.340mmol/L(2.39~5.72mg/dL) |
Reference ranges may vary between laboratories. Every laboratory should establish its own reference range as needed.
Lysun Profile
Hangzhou Lysun Biotechnology Co., Ltd. was established in 2018. Our company is located in the beautiful city of Hangzhou Binjiang High-tech Zone. Our mission is to provide people with the most convenient,accurate and affordable products to prevent diseases.
Lysun Certificate
Lysun Equipment
FAQ
Q1: Are you a manufacturer or a trade company?
A: We are manufacturer.
Q2: Could you print our company LOGO on the package?
A: Yes, we can do that.